Nano Spectrophotometer FM-NS-A100

SKU: 9e1e7dfe1074 Categories: , Brand:

Nano Spectrophotometer FM-NS-A100 is intended for quick, precise, and sensitive measurements of biological samples. Equipped with a 7-inch touchscreen and an android operating system, this device offers an intuitive user interface. A variety of innovative features assist laboratory procedures. It is ideal for both absorbance and fluorescence measurements in labs requiring high precision and reliability while using a minimal sample volume. The tool offers extensive capabilities and robust performance for a variety of laboratory applications.

Maximize Your Machine Potential

At Dyrec, we are committed to empowering your business with the latest solutions. Our top-quality rotary joints are designed to optimize the performance of your equipment.

Take the next step toward success by contacting us or downloading our comprehensive product guide. Your journey toward success begins here.

Description

Features

  • Nano Spectrophotometer has user-friendly interface
  • Quick detection within 2 to 6 seconds
  • Automatic detection function
  • Long shelf life of xenon flash lamp
  • Excellent repeatability
  • Built-in printer facilitates immediate printout of results
  • Exportable results

Applications

Nano Spectrophotometer is widely used across molecular biology, food, environmental sciences, forensic sciences, pharmaceuticals, and clinical diagnostics.

Absorbance precision0.002 Abs
Wavelength detector2048 element linear CMOS array
Sample volume0.5μl to 2μl (recommended 2μl)
Absorbance accuracy±1% (7.332 at 260nm)
Power adaptorDC 12V 4A
Wavelength range180 to 910 nm
Measurement time<6 seconds
Power consumption48W
Wavelength accuracy±1 nm
Path lengths0.03 mm, 0.05 mm, 0.1 mm, 0.2 mm, 1 mm
Spectral resolution≤ 1.5 nm (FWHM at Hg 253.7 nm)
Light source / lifeXenon / > 109 flashes
Absorbance range0.02 to 750 (260 nm, equivalent to 10 mm wavelength)
Concentration range2 ng/μl dsDNA to 37500 ng/μl dsDNA
OD600 absorbance range0 to 4.000 Abs
Fluorometer sample volume1 to 20 μl
Fluorometer measurement time3 seconds
Absorbance stability[0,3) ≤ 0.5%, [3,4) ≤ 1.5%
Fluorometer concentration range0.01 ng/μl to 120 ng/μl dsDNA HS, 0.2 ng/μl to 2000 ng/μl dsDNA BR, 0.05 ng/μl to 240 ng/μl Oligo, 0.1 mg/ml to 20 mg/ml protein BR
Absorbance repeatability[0,3) ≤ 0.5%, [3,4) ≤ 1.5%
Repeatability< 1.5%
Linear dynamic rangeR2 ≥ 0.995
Light sourceLED
HeatingUp to 37°C
Excitation wavelengths470/625 nm
Stirring speed100 to 900 rpm
Emission wavelengths525/690 nm
Dimensions (W×D×H)300 × 232 × 196 mm
Packing dimensions (W×D×H)450 × 400 × 350 mm
Net weight3.5 Kg
Gross weight7 Kg