Atomic Absorption Spectrophotometer FM-AAS-A101

SKU: 3a3ebb35b80d Categories: , Brand:

Atomic Absorption Spectrophotometer FM-AAS-A101 is a microprocessor controlled analyzer based on analysis of metal ions by flame and graphite furnace atomizer with absorption at 190 to 900 nm wavelength. Czerny-Turner monochromator 1800 line/mm grating optics and variance of bandwidth provides user flexibility. Provided with deuterium hollow cathode lamp, excludes interference of molecular absorption and Titanium Burner increases analysis, efficiency and precision of flame. Its double beam radiation sources enable good basic line stability. Graphic analysis system provides data on analysis of entire process i.e. measured value, temperature, procedures, time, etc.

Maximize Your Machine Potential

At Dyrec, we are committed to empowering your business with the latest solutions. Our top-quality rotary joints are designed to optimize the performance of your equipment.

Take the next step toward success by contacting us or downloading our comprehensive product guide. Your journey toward success begins here.

Description

Features

  • Integrated flame/graphite furnace atomization system, changeable with flame emission burner
  • Turret for 8 pre-aligned Hollow Cathode Lamps fitted with pre-heating, and individual pre-centered lamps
  • Spectral bandwidth of 0.1, 0.2, 0.4, 1.0 and 2.0 nm
  • Characteristic concentration of copper: 0.02 μg / ml / 1 %
  • Detection limit of copper: 0.004 μg / ml
  • PID and Dual curve mode light-controlled temperature control technique ensures fast heating and analytical sensitivity
  • Fully automated wavelength scanning and peak searching
  • Advance software for complete parameter settings
  • Reliable fully automatic graphite furnace analysis
  • Alarm feature to protect fuel gas leakage, abnormal flow, insufficient air pressure and abnormal flame extinction in flame system
  • Analysis of elements by flame absorption method and emission method
  • Strong database system which possesses more than 500 data self-storage
  • Results printed as parameter, data result and diagram

Applications

It is suitable to analyze trace amount of element from mixture, also in forensics and chemistry, biological science to detect toxic elements from blood and drug samples.

Wavelength&range&190 to 900 nm
Wavelength&accuracy± 0.15 nm
Wavelength&repeatability≤ 0.04 nm
Spectrum&bandwidth0.1, 0.2, 0.4, 1.0 and 2.0 nm
Base&line&stability≤ 0.002 Abs / 30 min
Grating1800 lines / mm
InflamerAll-metal titanium burner
AtomizerEfficient glass atomizer
LampHollow cathode lamp
Lamp&stand8 lamps automatic turret, automatic alignment
Characteristic&concentration&of&copper0.02 μg / ml / 1 %
Detection&limit&of&copper0.004 μg / ml
D2 Background&calibration&ability1 A ≤ 30 times correction ability
Measurement Repeatability0.005
Flame systemAir-acetylene flame system
Acetylene air burner100 mm
Ignition dynamic baseline drift≤ 0.006 Abs / 30 min
Characteristic viscosity≤ 0.025 μg / ml / 1 %
Standard deviation of the accuracy≤ 0.5 % (Cu, absorbance > 0.8 A) (detection limit Cu ≤ 0.008 μg / ml)
Protection systemCan inevitably cut off gas if pressure is low, power cut-off, flame out and alteration of the burner
Power&supply220 V ± 22 V AC
Dimension700 × 550 × 450 mm
Net weight75 kg

Produk Lainnya