Description
Features
- Integrated flame/graphite furnace atomization system, changeable with flame emission burner
- Turret for 6 pre-aligned hollow cathode lamps fitted with pre-heating, and individual pre-centered lamps
- Spectral bandwidth of 0.2 nm, 0.4 nm, 0.7 nm, 1.4 nm, 2.4 nm, 5.0 nm
- Characteristic concentration of copper: ≤ 0.04 mg/ml/1%
- Detection limit of copper: ≤0.008 mg/ml
- Built-in computer data processing system and LCD display
- Double-beam system compensates light source and wavelength drift caused by temperature variation
- Gas path system equipped with precision pressure stabilizing and current stabilizing devices
- Multi-functional analysis mode: flame absorption, flame emission, graphite furnace atomic absorption and hydride generation
- Integrated functions: integral holding, peak height and area, auto zero adjusting, deuterium lamp background correction, multi-linear and nonlinear curves fitting
- Various parameters and working curves displayed in screen and report printing
- It is equipped with autosampler and RS232 interface for externally linking PC
Applications
Used in different areas of chemistry such as clinical analysis of metals in biological fluids and tissues such as whole blood, plasma, urine, saliva, brain tissue, liver, hair, muscle tissue, semen, in some pharmaceutical manufacturing processes, minute quantities of a catalyst that remain in the final drug product, and analyzing water for its metal content.