Oil Particle Counter FM-OPC-A100

SKU: 944057a7633d Category: Brand:

Oil Particle Counter FM-OPC-A100 is adopted with counting principle of the photoresist (shading) method and high-pressure syringe pump sampling method, is designed as per ISO11171-1999 standard configurations. With positive and negative pressure combined sample injection system and 16 detection channels, features 0.8 to 600 µm of particle size range with 0.2 to 1000 ml sampling volume and 12000 to 40000 particles/ml of maximum particle concentration. Designed with high-precision laser sensor, built-in multiple calibration curves, single-channel and multi-channel calibration, offers full-function automatic operation through touch screen display. Semi-conductor laser as light source, built-in data analysis system, built-in air purification system with pressure sensor, and mass data storage, it supports RS232 interface and U-disk storage.

Maximize Your Machine Potential

At Dyrec, we are committed to empowering your business with the latest solutions. Our top-quality rotary joints are designed to optimize the performance of your equipment.

Take the next step toward success by contacting us or downloading our comprehensive product guide. Your journey toward success begins here.

Description

Features

  • Adopted with counting principle of the photoresist (shading) method
  • High-pressure syringe pump sampling method, fast and convenient sampling
  • Designed as per ISO11171-1999 standard configurations
  • Positive and negative pressure combined sample injection system
  • 16 detection channels, ensures efficient oil contamination analysis
  • 0.8 to 600 µm of particle size range and 0.2 to 1000 ml of sampling volume
  • 12000 to 40000 particles/ml of maximum particle concentration
  • High-precision laser sensor, offers wide test range with stable performance
  • Single-channel and multi-channel calibration with built-in multiple calibration curves
  • Full-function automatic operation through touch screen display
  • Semi-conductor laser as light source, low noise and high resolution
  • Built-in data analysis system with mass data storage
  • Built-in air purification system with pressure sensor, ensures safe operation
  • Supports RS232 interface and U-disk storage
  • High-efficient, stable and reliable unit with easy operation and data processing

Applications

Oil Particle Counter is used across the field of insoluble particles for organic liquid and polymer solutions, pollution degree of oil, organic liquid and polymer solution etc.

Particle size range0.8 to 600 µm
Detection channel16
Sensitivity0.8 µm or 3 µm
Resolution< 10%
RepeatabilityRSD < 2%
Light sourceSemiconductor laser
Sample detection viscosity≤ 650 cSt
Sampling volume0.2 to 1000 ml (interval: 0.1 ml)
Sampling accuracyBetter than ± 0.5%
Sampling speed5 to 80 l/min
Air pressure chamberPositive and negative air pressure chamber device
Maximum vacuum of air pressure chamber-0.08 Mpa
Maximum positive pressure of air pressure chamber0.8 MPa
Maximum particle concentration12000 to 40000 particles/ml
Power supply110 to 245 V AC 50 Hz
Power70 W
Dimension (W×D×H)340×410×650 mm
Packaging dimension (W×D×H)850×560×700 mm
Net weight25 kg
Gross weight45 kg